GE Healthcare · MRSigna22247252224580

Prescan Failure due to 8Ch CTL Coil Cable.Auto and Manual Prescan Fails.Replace Coil cable wqr.Signa HDxT.2224725 Low si…

Prescan Failure due to 8Ch CTL Coil Cable.Auto and Manual Prescan Fails.Replace Coil cable wqr.Signa HDxT.2224725 Low signal detected.2417163 ASSY,FRU,CABLE.15.0

Symptom

Auto and Manual Prescan Fails

Error message

2224725 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan) See Long Description Error: 2224725 Host: hopgemr1a Process: NSP File: APS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan).

2224580 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224580 Host: hopgemr1a Process: NSP File: APS:aps_cf Line: 11 Auto Prescan failed. Signal was too small to process - use Manual Prescan

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2417163 ASSY,FRU,CABLE,VALUE 8CH CTL 2417163, ASSY,FRU,CABLE,VALUE 8CH CTL
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues