GE Healthcare · MRExcite2253850
Collapsed Image and Gradient Link hammer Test Failures Due to loose gradient data fiber.MRIQ Collapsed Image Bright Whit…
Collapsed Image and Gradient Link hammer Test Failures Due to loose gradient data fiber.MRIQ Collapsed Image Bright White Line Down or Across Middle of The Image.Seated the "% GRAD" Fiber Optic Cable…
Symptom
MRIQ Collapsed Image Bright White Line Down or Across Middle of The Image
Error message
2253850 Gradient Link Hammer test failed. The GP received an invalid %c axis data value via the gradient data link. Check the gradient data and clock fiber optic cable connections 2253850 Gradient Link Hammer test failed. The GP received an invalid %c axis data value via the gradient data link. Check the gradient data and clock fiber optic cable connections
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.