Pular para o conteúdo principal
GE Healthcare · MROptima2267801

Unable to Scan with Message The IRF3 is not ready.due to Fiber Optic Transceiver in %-axis Port of XGA Control Bd in Cro…

Unable to Scan with Message The IRF3 is not ready.due to Fiber Optic Transceiver in %-axis Port of XGA Control Bd in Cronus chassis.

Symptom

Unable to Scan with Message The IRF3 is not ready. See Long Description Customer is unable to scan due to message in message window located at bottom of Scan Desktop which states “The IRF3 is not ready.->”

Error message

2267801 Safe mode startup of the XGD (XGA-SB Y axis) failed. Check EMO reset and perform a TPS reset. See Long Description 2267801 "Safe mode startup of the XGD (XGA-SB Y axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative.->"

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5183329-2 Fiber Optic Transceiver SFP 4Gb Multimode, RoHS-Compliant Note, as of March 24, 2010 this part can not be ordered yet. It will be a FRU. If you cant order the transceiver order the higher FRU.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues