2267863 The X axis XGA is reporting a I_Coil Wire Fault due to XGA amplifier.

Symptom

Cannot scan with Error: 2267863 The % axis XGA is reporting a I_Coil Wire Fault Cannot scan with Error: 2267863 The % axis XGA is reporting a I_Coil Wire Fault.

Error message

2267863 The %axis XGA is reporting a I_Coil Wire Fault See Long Description Extended Error: The cables between the coil current sensor and the SB board is loose or has fallen off. Ensure cables J5 and J6 of the XGA are connected to the corresponding axis current sensor at the top of the Power Cabinet.

2267778 An error on the XGD has prevented TPS reset from completing Successfully Error: 2267778 Host: UCSF-OIMR1 Process: NSPFile: SCP:GradientHwo Line: 1986Exception Class: Unknown Severity: UnknownFunction: Gradient: N/APSD: Unknown Coil: Unknown Scan: -1/-1 An error on the XGD has prevented TPS reset from completing successfully. Please review the log for XGD faults to determine the cause and action(s) to take. Perform another TPS reset and if this error occurs again, contact your service representative.

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
Area
MR