Coil ID Fails due to Neck Chest Unit.

Symptom

Coil ID not recognized intermittently. Only one coil is affected.

Error message

2273003 Coil validation error. An incorrect CRC was detected in the coil ID data from the coil. The coil is programmed incorrectly. See Long Desc. 2273003 Coil validation error. An incorrect CRC was detected in the coil ID data from the coil. The coil is programmed incorrectly. Connector: %b Extended Error: If this happens whenever the coil is plugged in, this is most likely a bad coil. If not, this could indicate a bad connection at the coil interface, a bad coil, bad cables to the RF Hub RFCB, or a bad RFCB. Try connecting another coil. Run the Coil ID Diagnostic or Coil Datapath Diagnostic to troubleshoot cabling issues and to isolate the coil interface from the coil.

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Component
2423193 NECK CHEST UNIT 1.5T BRAIN SPINE FRU
Area
MR

Full cause & step-by-step resolution