Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed26010009350164

BrightSpeed HPower, Delight, Velocity or MERI HPower

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260100093 PDU HVDC Buss Fault Status. The HVDC did not charge correctly in the alotted time Mon Sep 5 13:55:52 2005 Error: 260100093 Host: tgp Process: tgp_subsys File: scIsrService.c Line: 246 Running diagnostic: kV and mA Test: Xray Functional Test Mon Sep 5 13:55:51 2005 Host : Table Gantry Processor Ermes # : 260100093 Exception Class : Abort Severity : Primary File : scIsrService.c Line# : 246 Function : SYSTEM MONITORING : HVDC Monitor Scan Type : Static Scan: 50164/1/1 PDU HVDC Buss Fault Status. The HVDC did not charge correctly in the alotted time. Threshold 1: > 200V in 200ms Threshold 2: > 500V in 1.2s. Possible causes: HVDC Cable Short Circuit, SlipRing, or Jedi Inverter.

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2387385 Power Pan Asm. 550V See Long description. 2387385 Power Pan Asm. 550V Part #: 2387385 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: POWER PAN ASM 550V FRU: Yes Status: Orderable Parts Offered as: Current: 2387385 Harvested: Not Available Exchange: Not Available Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.