Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT23000006929144370005000324400036000

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Do Not Use - use one of the other Error: 230000069 Two, or more, unexpected events were detected during scan data transfer. The 230000069 Error has several varation see below. Sun Sep 5 11:10:31 2004 Error: 230000069 Host: darc Process: dataacq File: dip_thread.cxx Line: 396 Function: Data Acquisition : Scan Data Acquisition Scan Type: Scout Scan: 4167/1/1 Scan Seq Id: 1000 Two, or more, unexpected events were detected during scan data transfer. Bit Count Event description 3 1 TAXI modem violation 5 1 View length error Mon Dec 15 10:40:11 2008 span style="COLOR: #ff0000"> Error: 230000069 Host: darc Process: dataacq File: hddip.cxx Line: 720 Function: Data Acquisition : Scan Data Acquisition Scan Type: Helical Scan: 29144/9/3 Scan Seq Id: 37000 Two, or more, unexpected events were detected during scan data transfer. Bit Count Event description 10 1 View Data Corruption Detected: ---------------------- Bad View Details: Bit Count Event description 0 1 Checksum Field Error 1 1 View Number Field Error 2 1 Row Number Field Error 3 1 Data Length Field Error 4 1 Data Type Field Error 5 1 Magic Number Field Error 7 1 Consecutive Corrupted Segments Detected Mon Dec 22 07:16:59 2008 Error: 

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2285086-2 400MB RX, HSDCD SLPR 2285086-2 400MB RX, HSDCD SLPR See Long Description for Details. 2285086-2 400MB RX, HSDCD SLPR (replaces 2285086) Part #: 2285086-2 Part Type: GP CONSUMABLES Part Weight: 1.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 400MB RX, HSDCD SLPR, ESD PROTECTION FRU: Yes Status: Orderable Parts Offered as: Current: 2285086-2 Harvested: Not Available Exchange: Not Available Comments: FLS - HSDCD 400MB RX SLIPRING. PROVIDED BY SCHLIEFRING. REPLACES P/N 2285086. USE CURRENT STOCK FIRST. ADDED PER PCN 220321 BY MS20544, CSV 12/5/01 ONE USE IS QXI LIGHTSPEED DO NOT DISCOUNT - PER SHAKUN
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.