Pular para o conteúdo principal
GE Healthcare · MRExciteOvationSigna22776512223859227767215373572223883

TPS Reset Fails with VRE Errors Due to Sun VRE.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2277651 ICN Hard Disk Health Diagnostic failed. Mon Feb 23 06:44:11 2009 Error: 2277651 Host: EPMR02 Process: diagCoord File: icnDiskHealth.pl Line: 363 Exception Class: Notification Function: Software/Firmware/Config: N/A ICN Hard Disk Health Diagnostic failed. icn1 - No disks to process->

2223859 The TPS Reset Failed because of a VRE Reset Failure See long description. 2223859 Host: ghmr002 Process: NSP File: TardisReset.cc Line: 469 Exception Class: Unknown Function: Recon: Data Acquisition The TPS Reset failed because of a VRE reset failure. Extended Error: This indicates that a failure to reset the VRE properly caused the TPS Reset to fail. Please refer to the VRE Troubleshooting Guide for further troubleshooting this issue, which can be found in the MR Service Documentation.

2277672 TPS Reset ICN Boot Log Analysis detected an improper shutdown Mon Feb 23 06:44:04 2009 Error: 2277672 Host: EPMR02 Process: diagCoord File: icnBootLogCheck.pl Line: 358 Exception Class:Notification Function: Software/Firmware/Config:N/A TPS Reset (ICN Boot Log Analysis) detected an improper shutdown on ICN:[icn1 MAC=00:30:48:5B:D5:22 SERIAL=1537357].->

2223883 VRE reset failed due to failure to

Applies to

Models
Excite, Ovation, Signa
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Component
5130108 Value Image Compute Node Value ICN used on HDe, ProfileHD and OvationHD
Area
MR

Full cause & step-by-step resolution