GE Healthcare · MRSigna2244847
Error 2244847 The GP reported a Gradient Amplifier Y Axis Internal Wiring Fault or Control Undervoltage due to STIF card…
Error 2244847 The GP reported a Gradient Amplifier Y Axis Internal Wiring Fault or Control Undervoltage due to STIF card in MGD chassis.
Symptom
Will Not Scan with Gradient Errors.
Error message
2244847 The GP reported a Gradient Amplifier % Axis Internal Wiring Fault or Control Undervoltage Error: 2244847 Host: SEMRADMR2 Process: NSP File: SCP:Grad Line: 2615 Exception Class: Unknown Severity: Unknown The GP reported a Gradient Amplifier X Axis Internal Wiring Fault or Control Undervoltage.
Applies to
- Models
- Signa
- Product
- 3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.