Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260007370

LightSpeed 4.X - GRE

Symptom

Customer sees a Pop-up window indicating the DAS temperatures are low. This does not prevent scanning only a warning. Customer sees a Pop-up window indicating the DAS temperatures are low. This does not prevent scanning only a warning

Error message

Error: 260007370, DAS Temperature is LOW. See Long Description for complete error Mon Feb 20 07:43:59 2006 Error: 260007370 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 490 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:43:59:965 Log Series: 32 DAS Temperature is LOW. Lowest temperature was seen is on Converter Board 31. Measured: 10.5 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. See Following Error Message for Converter Card List

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2294856-2 (old # 2294856) DHCB assembly Detector Heater Control Board See LD for femc info and picture 2294856-2 (old # 2294856 ) DHCB assembly Comments: FLS - DHCB, BOARD / COVER ASM (H3) DETECTOR HEATER CONTROLLER BD picture Part #: 2294856-2 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: DHCB, Board-Cover Asm EMC ED2 FRU: Yes Status: Orderable Parts Offered as: Current: 2294856-2 Harvested: Not Available Exchange: Not Available Comments: Added per EMC2 RSPL, this part is EMC2 compliant replacement for 2294856. ms13452 050509
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues