Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1111710008

LightSpeed RT16 & RT16 Xtra and Vision

Symptom

CU does not want the Dose Structured Report, Series 997, to be generated on their exams. See long description for details CU does not want the Dose Structured Report, Series 997, to be generated on their exams. What is a dose structured report ? It is a structured report object, which complies with DICOM Standard 3.16 for dose reporting. The DICOM fields in the report can be manually exported to a data base to track patient dose.

Error message

Error: 11117 Remote does not support push for SOP_Class_UID See Long Description Two version of this error Exist. “DICOM Enhanced SR SOP” class. UID 1.2.840.10008.5.1.4.1.1 .88.22 “DICOM X-ray Radiation Dose SR SOP” class UID 1.2.840.10008.5.1.4.1.1 . 88.67 Fri Feb 13 15:22:35 2009 Error: 11117 Host: msec8x1 Process: dcserver File: dcspush.c Line: 2456 Remote does not support push for SOP_Class_UID 1.2.840.10008.5.1.4.1.1.88.67 , so all images with this SOP_class_UID have been rejected Fri Feb 13 15:22:35 2009 Error: 11117 Host: msec8x1 Process: dcserver File: dcspush.c Line: 2456 Remote does not support push for SOP_Class_UID 1.2.840.10008.5.1.4.1.1.88.22 , so all images with this SOP_class_UID have been rejected “DICOM Enhanced SR SOP” class. UID 1.2.840.10008.5.1.4.1.1 .88.22 “DICOM X-ray Radiation Dose SR SOP” class UID 1.2.840.10008.5.1.4.1.1 . 88.67 * SR stand for "Structured Report"

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
10BW27.7 Software for BSD Delight (AIO Console) and RT16 Vision See Long Description 10BW27.7 Software for BSD Delight (AIO Console) and RT16 Vision This software was released via FMI 22926 FMI Title: 10BW27.7 Software Release for BSD Delight and RT16 Vision Furnished Materials FMI22926 Kit (Part Number: 5407235): Items Part Number Part Description 5407235-1EN FMI 22926 Instruction (Hard Copy) 5408645 10BW27.7 jewel case 5405943 OS 6.3.8 (DVD) 5405944 App (DVD) 5407148 SOL (CD) 5407135 SP1.0 for 10BW27.7 (CD) 5390977-1EN SP2.0 Common Addendum TWA_English (Hard Copy) 5390977-299 SP2.0 Common Addendum TWA_Muli-language (CD) • 5405945- Class C (Tab #182) • 5405946- Class M (Tab #182)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.