GE Healthcare · MRDiscovery222755622678362267778
2267836 The Z axis XGA is reporting a cold plate over-temperature The cold-plate temperature is 59C due to XGA 5184566-2…
2267836 The Z axis XGA is reporting a cold plate over-temperature The cold-plate temperature is 59C due to XGA 5184566-2.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2227556 *** Reset/Download TPS was not successful. *** 2227556 sdm sdm_main.c 320 *** Reset/Download TPS was not successful. *** 2267836 The Z axis XGA is reporting a cold plate over-temperature The cold-plate temperature is 59C. see long Error: 2267836 Host: DVMR Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The Z axis XGA is reporting a cold plate over-temperature. The cold-plate temperature is 59C.-> 2267778 An error on the XGD has prevented TPS reset from completing Successfully Error: 2267778 Host: UCSF-OIMR1 Process: NSPFile: SCP:GradientHwo Line: 1986Exception Class: Unknown Severity: UnknownFunction: Gradient: N/APSD: Unknown Coil: Unknown Scan: -1/-1 An error on the XGD has prevented TPS reset from completing successfully. Please review the log for XGD faults to determine the cause and action(s) to take. Perform another TPS reset and if this error occurs again, contact your service representative.
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- 20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
- Component
- 5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.