Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT18280450080

LightSpeed 2.X (Plus, QX/i) - GRE

Symptom

Gantry Display blank. Gantry Display blank

Error message

Error: 182804 The Gantry Display is NOT Initialized Properly See the LD for the full error code Error: 182804 The Gantry Display is NOT Initialized Properly Wed Feb 8 07:20:05 2006 Error: 182804 Host: ETC Process: Scan File: etc_in_mail.c 1.38 4/16/01 Line: 648 Function: No System Function Reported Scan: 50080/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 07:20:05:645 Log Series: 37 HW/SW error: Communication Problems with the Gantry Display. The Gantry Display is NOT Initialized Properly. (Scanning is NOT allowed if the Gantry Display is NOT Working)

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2256730-3 Display assembly used on 2.X/3X/4X5X/7X See long description for details. Part #: 2256730-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: H-POWER COMPATIBLE GANTRY DISPLAY FRU: Yes Status: Orderable Comments: FLS - H-POWER COMPATIBLE GANTRY U/O CT H POWER DISPLAY
Area
CT

Full cause & step-by-step resolution