Unable to Scan with RF Air Temperature Faults Due To Customer Coolant Chiller. unable to scan with HEC errors.

Symptom

Unable to Scan with HEC Errors To toggle the pumps and blower On and Off, do the following: Blower: Press and hold the Up arrow ( <^> ) and press <F1> . Gradient Coil Pump: Press and hold the Up arrow ( <^> ) and press <F2> . Power Electronics: Press and hold the Up arrow ( <^> ) and press <F3> .

Error message

2268905 RF air temperature fault detected. Scanning will be inhibited until the RF air temperature matches expected. See Long Description 2268905 NSP SCP:SriCanHWO 1998 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 RF air temperature fault detected. Scanning will be inhibited until the RF air temperature matches expected Temperature difference: 107 0.1 degrees C Tolerance: 30 0.1 degrees C-> Extended Error: The Heat Exchanger (HEC) manages RF air temperature. This message indicates that the HEC was unable to achieve the requested RF air temperature.

2268901 RF coil airflow fault detected. See Long Description Scanning will be inhibited until the airflow is as requested Airflow difference: 117 fpm Tolerance: 100 fpm-

2261733 Heat Exchanger Cabinet (HEC) detected customer coolant temperature is out of specification (too high). See Long Description. 2261733 NSP SCP:CoolingHWO 2836 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Heat Exchanger Cabinet (HEC) detected customer coolant temperature is out of specification (too high). Pl

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.