GE Healthcare · MROptima2267446222755622569252267798
Fails TPS Reset when booting up to Applications or when Resetting TPS due to Loose Connection at X-Axis XGA-J12.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2267446 The XGD module did not send a Fiber Link Init Complete 2227556 *** Reset/Download TPS was not successful. *** 2227556 sdm sdm_main.c 320 *** Reset/Download TPS was not successful. *** 2256925 StartupCondition error on SCP resulted in TPS reset failure. See Long Description 2256925 MOI_MR LxMxConsumer services/initialization/ServerInitManager 639 Server Name: SC StartupCondition error on SCP resulted in TPS reset failure. 2256925 StartupCondition error on Irf3 resulted in TPS reset failure. 2256925 Host: nbomr2 Process: LxMxConsumer File: services/initialization/ServerInitManager Line: 767Server Name: PS ^C StartupCondition error on Irf3 resulted in TPS reset failure 2256925 StartupCondition error on AGP resulted in TPS reset failure. See Long Description 2256925 MOI_MR LxMxConsumer services/initialization/ServerInitManager 639 Server Name: SC StartupCondition error on AGP resulted in TPS reset failure. 2256925 StartupCondition error on XGD StartupCondition resulted in TPS reset failure See Long Description 2256925 SLMAINMR LxMxConsumer services/initialization/ServerInitManager 767 Server Name: SC StartupCondition error on XGD StartupCondition resulted in TPS reset fail
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.