Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT23002303323001708850019

LightSpeed 4.X - Octane

Symptom

Fastcal fails the mylar window check. Fastcal fails the mylar window check

Error message

Error: 230023033 Mylar Window Check has failed. Mylar Window may be dirty. Calibration vectors generated with the dirty mylar window may cause Image artifacts. See Long Description for complete error Sat Jan 8 08:06:05 2005 230023033 Host: uhct Process: dailyPrepRx File: DailyPrepUtils.c Line: 2967 Mylar Window Check has failed. Mylar Window may be dirty. Calibration vectors generated with the dirty mylar window may cause Image artifacts.

Error: 230017088 The mylar window check failed See Long Description for complete error Mon Aug 23 05:50:31 2004 Error: 230017088 Host: usafaultra Process: accs File: aircal processing.c Line: 1025 The mylar window check failed - Scan 50019/1/3

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.