Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT200109012200502003

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

HP Host computer intermittently locks up or will not boot up to applications HP 8000 or 8200 computer intermittently locks up or will not boot up to applications

Error message

Error: 200109012 Image Allocation Failed. Cannot build scan request. See Long Description for complete error Wed Mar 2 18:39:33 2005 Error: 200109012 Host: ct01 Process: scanRx File: UIRxToScanner.cxx Line: 1307 Function: Data Acquisition : OC Processing Image Allocation Failed. Cannot build scan request.

Error: 200502003, Cannot allocate image space for XXX images, allocated XX images. See LD for complete error Mon Feb 13 16:36:28 2006 Error: 200502003 Host: ls16dn Process: retroRecon File: srx_sma_manage_image_space.c Line: 690 Function: No system function reported Cannot allocate image space for 148 images, allocated 7 images.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2368697-3 HEWLETT-PACKARD HP8000 Computer used in Global Recon Engine (GRE) CONSOLE 2368697-3 GRE CONSOLE HP8000 2368697-3 HP8000 COMPUTER USED IN GRE CONSOLE Part #: 2368697-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: LINUX DDR WORKST'N PS2 MOUSE--100 PC FRU: Yes Status: Orderable Comments: 3/4/05 (212001261 Shawn Fox) HOST COMPUTER (HP). THISIS THE GOC2 HP 8000 COMPUTER. ***EXTREMELY IMPORTANT: NEW SOFTWARE, S/W LOAD PROCEDURE AND S/W OPTION KEYS ARE REQUIRED IF THE HP8000 COMPUTER IS REPLACED. Please contact the OLC for assistance with new Option Keys
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.