Pular para o conteúdo principal
GE Healthcare · MRSigna22247092224580

No Receive Signal with any Multicoils Due To No Preamp Bias Voltage Due to Shorted White Receive Cable.

Symptom

No Receive Signal with any Multicoils

Error message

2224709 Low signal detected. Please check the signal level for receiver x or receiver x+1 while using Manual Prescan. Error: 2224709 Host: NMI_Langhorn Process: NSP File: TPS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver 1 (or receiver 2 while using Manual Prescan). Note that for 3 coil scans, one coil will cause a low signal detected message to occur. Error: 2224709 Host: NMI_Langhorn Process: NSP File: TPS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver (X + 1) while using Manual Prescan). Note that for 3 coil scans, one coil will cause a low signal detected message to occur.

2224580 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224580 Host: hopgemr1a Process: NSP File: APS:aps_cf Line: 11 Auto Prescan failed. Signal was too small to process - use Manual Prescan

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
46-317220G20 MULTI-COIL CABLE TAKE-UP RF CABLES 46-317220G20: MULTI-COIL CABLE TAKE-UP RF CABLES. Cables and track for GE S-series long bore magnet.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required