GE Healthcare · CT26000232313501
Lightspeed 1.X - Octane
Symptom
Scan aborts. Scans abort
Error message
Error: 260002323 Acquisition hardware (DCB) has detected an Unacceptable Amount of Acquisition Trigger Jitter See long description for full error Wed Aug 6 13:55:53 2008 Error: 260002323 Host: DAS Process: DCB Firmware File: abort.c Line: 315 Function: DATA ACQUSITION : Scan Data Trigger Generation Scan: 13501/6/1 Type: Helical Exception Level: Sec/Soft Time: 13:55:52:965 Log Series: 281 Scan ABORT -- DAS Has Detected An Error.^B Acquisition hardware (DCB) has detected an Unacceptable Amount of Acquisition Trigger Jitter. (Trigger Frequency: 1230 hertz). (State: Normal State Command Level: Scan in Progress ScanState: Collecting Scan Data State
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2254362 SIGNAL BRUSH MODULE_FOR HSDCD DRUM - used on H1 gantry See Long Description for part details Part #: 2254362 Part Type: GP CONSUMABLES Part Weight: 0.20 KGM Repairable: No Returnable: No Hazardous Material: N Description: SIGNAL BRUSH MODULE_FOR HSDCD DRUM FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required