LightSpeed 5.X (80/100kW)
Symptom
Scanner fails to boot to applications, Unable to verify DARC system startup. The system will not boot up to applications with a pink attention box: Unable to verify system functionality on startup. Please power cycle the console. If the system fails to startup after the power cycle, then please call GE Service. FE or OLE sees that they can telnet to the DARC but they cannot ping or rsh with the DARC. If the DARC is observed while booting, Drive X adapter X predicting future failure Press any key to continue is seen.
Error message
Error: 245047 Unable to allocate scan files See the long description for the full error code Thu May 12 17:43:37 2005 Error: 245047 Host: rrmcct16 Process: scanmgr File: scan_sequence.cxx Line: 1461 Function: System Control : Scan Control Scan Type: Scout Scan: 0/0/0 Scan Seq Id: 601000 Unable to allocate scan files /dev/md0: not enough valid hard drives - bus0(#), bus1(#) /dev/md0: unable to create a new disk array See long description for details For GRE systems with DARC I or DARC II When logged into the DARC as root and trying to query the SDDA Scan Disks, you get the errors below: cd /usr/g/bin ./gre-raid –q /dev/md0: not enough valid hard drives - bus0(1), bus1(0) /dev/md0: unable to create a new disk array When attempting to create a new disk array partition it also failed with not enough drives found: logged in as root rsh darc cd /usr/g/bin ./gre-raid –c -z The error messages below indicate a bad SDDA disk drive: /dev/md0: not enough valid hard drives - bus0(1), bus1(0) /dev/md0: unable to create a new disk array RAID-0 active with (2) drives, 256k chunk size, and 69GB capacity 35.6MB/sec...min performance failed. Note that bus0 sees the drive but bus1 does not! On a VCT
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5114772 Darc2 Computer Dual Disk used ONLY on 16 slice MDAS and 4/8/16 slice GDAS See LD for more info 5114772 Darc2 Computer Dual Disk used ONLY on 16 slice MDAS and 4/8/16 slice GDAS Part #: 5114772 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: DARC Computer - Dual Disk H16 MDAS, GDAS FRU: Yes Comments: Added per GOC4 RSPL Replace the DARC2: 1) If you get a Westville DARC P/N 5114772 load the DARC software to recover. 2) If you get Jarrell p/n 5147442 perform LFC using 06MW29.7 software.
- Area
- CT