Auto Prescan failed, RF Amplifier is not ready.

Symptom

Auto Prescan failed, RF Amplifier is not ready Getting another message that RF Amplifier is powering but message never goes away.

Error message

2254051 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. See Long Description Fri Aug 20 08:33:11 2004 Error: 2254051 Host: lx-mr Process: NSP File: SCP:PmMdsHWO Line: 1019 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 35 pulse width: 33 duty cycle: 09 Fault in head amplitude. This could also occur when power cycling SSM.

2224666 Auto Prescan Failed, RF Amplifier is Not Ready NOTE: When using this error to create new knowledge, please include any associated error codes. 2224666 is a generic error code which requires further information. Mon May 17 08:19:08 2004 Error: 2224666 Host: lx_mr Process: psc File: PscMainControl.cc Line: 1296 Auto Prescan failed, RF amplifier is not ready.

2248720 The RF Cabinet MDS peripheral received an incorrect response from the power monitor. Possibly a bad SSM APM Board, a bad SSM CPD Board, or a bad connection between them. It may be necessary to perform a TPS Reset.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues