Pular para o conteúdo principal
GE Healthcare · MR222548722199712219981

TCU Communication Failures Coil ID Fails To Work And Z2 Shim Supply Not Responding due to damaged cable Run 879.

Symptom

TCU Communication Failures Coil ID Fails To Work And Z2 Shim Supply Not Responding. HFO system running HFO2 s/w with an Octane II computer. Unable to communicate with the TCU, also coil id is not functioning. When they plug the head coil into the MSHS it stays "RED", indicating coil ID failed. The thing to note here is that on the OpenSpeed, the serial line for the coil ID and the serial line for the TCU and the Z2 shim supply goes out through the SCSI Serial Expansion box. Did an scsistat and find the system does not identify the SCSI expansion box. HFO system running HFO2 s/w with an Octane II computer. Unable to communicate with the TCU, also coil id is not functioning. When they plug the head coil into the MSHS it stays "RED", indicating coil ID failed. The thing to note here is that on the OpenSpeed, the serial line for the coil ID and the serial line for the TCU and the Z2 shim supply goes out through the SCSI Serial Expansion box. Did an scsistat and find the system does not identify the SCSI expansion box.

Error message

2225487 RFampType in MRconfig.cfg file does not match RF Amplifier Type determined by SPI. 0.07T See Long Description Mon Aug 10 06:24:53 2009 Error: 2225487 Host: SUNRGESGMR2 Process: NSP File: HPC_PORT_ID:spi_rf_rd_freq_handler Line: 3040 RFampType in MRconfig.cfg file does not match RF Amplifier Type determined by SPI. Please update MRconfig.cfg file with correct RFampType. Actual RF Amplifier Type: 0.07T Single Solid State Analogic Note: " 0.07T " appears to be a typo in this error message. This error was logged on an 0.7T (OpenSpeed) system, so the RF Amplifier Type is "0.7T", not "0.07T" (seven tenths of a Tesla, not seven hundredths of a Tesla).

2219971 Error in initializing TCUs. Could not initialize TCUs Mon Dec 1 10:58:34 2003 2219971 1 hvmr port_mgr PMMain.cc 447 Error in initializing TCUs. Could not initialize TCUs

2219981 Z2 supply is not up or not responding, Retrying Mon Dec 1 10:58:58 2003 2219981 1 hvmr port_mgr PMMain.cc 552 Z2 supply is not up or not responding, Retrying

Applies to

Product
HF OpenSpeed HFO1 HFO2 and HFO3 Only.
Software
HFO3
Component
2262695 Run 879 Coil ID cable
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues