Intermittent Warnings of high Patient Comfort Temperature Readings.Clean or Replace Blower Air Filter wqr.Signa LX 9.1.2…

Intermittent Warnings of high Patient Comfort Temperature Readings.Clean or Replace Blower Air Filter wqr.Signa LX 9.1.2225682 The patient comfort level is warmer than normal.2121199-5 Filter

Symptom

Intermittent Warnings of high Patient Comfort Temperature Readings Intermittent Warnings of high Patient Comfort Temperature Readings.

Error message

2225682 The patient comfort level is warmer than normal. Error: 2225682 Host: TRIOBMR1 Process: NSP File: HPC_PORT_ID:spi_bore_wall_temp_mon Line: 400 The patient comfort level is warmer than normal.^A The patient may be uncomfortable. Further increases in the temperature will inhibit scanning. Verify: 1. Patient fan is ON. 2. Room temperature is normal. 3. Air flow through the bore is unobstructed.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2121199-5 BRM Blower Cabinet Filter Part changes to 2279001 quanity of 10. Single filters can no longer be ordered. 2121199–5 - AIR FILTER, DIRECTION 2138247 Rev 14 page 2-100