Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed26011821310153

BrightSpeed Select, Delight Select or MERI Select

Symptom

Scan hardware aborts scanning and allows resumes with streak artifact on images See Long Description for details and example images Scan hardware aborts scanning and allows resumes with streak artifact on images.

Error message

Error: 260118213 (60-0316) Spit Max error. PPC board has detected the number of tube spits during exposure has reached the limit See Long Description for two error examples Sat Dec 17 18:32:18 2005 Error: 260118213 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Sat Dec 17 18:32:17 2005 Host : Jedi Ermes # : 260118213 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Axial Scan: 10153/2/2 (60-0316)Spit Max error. PPC board has detected the number of tube spits during exposure has reached the limit (see theory of operation, software section).^C Z6,Z7= ch,0h

Applies to

Models
BrightSpeed
Product
BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
D3146T SOLARIX 350 TUBE UNIT USED on BrightSpeed Select (lite) See long description for details Part #: D3146T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: REPLACEMENT SOLARIX 350 TUBE UNIT FRU: Yes Status: Orderable Comments: red tube, added per to BrightSpeed Lite RSPL by sso#305010201
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues