Symptom

"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors

Error message

Error: 260107251 Axial Motor Control (AMC) logic lost communication with the Axial Drive. See Long Discription for Details Error: 260107251 Axial Motor Control (AMC) logic lost communication with the Axial Drive. Mon Jul 19 07:41:43 2010 Error: 260107251 Host: tgp Process: tgp_subsys File: ds301.c Line: 828 Mon Jul 19 07:41:41 2010 Host : Table Gantry Processor Ermes # : 260107251 Exception Class : Secondary Severity : Sec/Soft File : ds301.c Line# : 828 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 . Axial Motor Control (AMC) logic lost communication with the AXIAL drive (Comm Watchdog Failure)

Error: 260107800 DS301 CAN packet transmission error See long description Tue Oct 26 03:43:43 2010 Error: 260107800 Host: tgp Process: tgp_subsys File: ds301.c Line: 1019 Tue Oct 26 03:43:42 2010 Host : Table Gantry Processor Ermes # : 260107800 Exception Class : Secondary Severity : Pri/Soft File : ds301.c Line# : 1019 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 . DS301 CAN packet transmission error. Number of send attempts: 3. Maximum allowed send attempts: 3.

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.