GE Healthcare · CTHiSpeedLightSpeed11027327041932
HiSpeed QX/i - Octane
Symptom
No exams listed in Image Works or examRx with Media error: Unrecovered data block read error reported in OC SYSLOG
Error message
Unrecovered data block read error in the /usr/adm/SYSLOG Unrecover data block read error in the /usr/adm/SYSLOG Nov 29 12:25:21 6A:anact03801 unix: dksc0d1: retrying request Nov 29 12:25:21 6A:anact03801 unix: dksc0d1: [Alert] Media error : Unrecovered data block read error (asc=0x11, asq=0x0) , (data byte 36) , Block # 1102732 (7041932) Unrecovered data block read error in the /usr/adm/SYSLOG Nov 29 12:25:21 6A:anact03801 unix: dksc0d1: <6> retrying request Nov 29 12:25:21 6A:anact03801 unix: dksc0d1: <6>[Alert] Media error<6>: Unrecovered data block read error (asc=0x11, asq=0x0)<6>, (data byte 36)<6>, Block #1102732<6> (7041932)<6>
Applies to
- Models
- HiSpeed, LightSpeed
- Product
- HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 101I.2_HSQXIM5 See LD for ClassM & C 101I.2_HSQXIM5 used on HiSpeed (YMS) only Class M software - 2395173 or 2401759 (6 pack) Class C software - 2395172 or 2401759 (6 pack) Patch G.6 Mandatory FMI# 25363 (P/N 5136941)
- Area
- CT
Full cause & step-by-step resolution