Pular para o conteúdo principal
GE Healthcare · CTHiSpeed182912

Gantry controls would not work when software was rebooted

Symptom

Gantry controls would not work when software was rebooted See the long description for the errors Thu Sep 16 06:48:08 2004 Error: 182904 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 370 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Unknown Exception Level: Pri/Most Severe Time: 06:48:00:070 Log Series: 20 A table side key is either stuck or pressed during initialization or just after Emergency Off released. Port value 0x0410. Disabling those keys detected being stuck Thu Sep 16 06:48:08 2004 Error: 182906 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 370 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 06:48:00:070 Log Series: 20 Gantry/Tableside Up Elevation button is non-functional. Use the Foot pedals. Thu Sep 16 06:48:08 2004 Error: 182912 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 406 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 06:48:00:075 Log Series: 20 Gantry/Tableside FAST button is non-functional

Error message

Error: 182912 Gantry/Tableside FAST button is non-functional See the LD for the full error code Error: 182912 Gantry/Tableside FAST button is non-functional Thu Sep 16 06:48:08 2004 Error: 182912 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 406 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 06:48:00:075 Log Series: 20 Gantry/Tableside FAST button is non-functional

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.