GE Healthcare · CTHiSpeed2006000071951020000043720000042323000906710989
HiSpeed QX/i - Octane
Symptom
System takes a scout but the tech cannot see it in ExamRx - document in RFS - critical issue System takes a scout but the tech cannot see it in ExamRx.
Error message
Error: 200600007, Could not get image space See Long Description for complete error Mon Feb 13 16:36:28 2006 Error: 200600007 Host: ls16dn Process: retroRecon File: UIRxToRetroRequest.cxx Line: 1402 Could not get image space for 19510/525/445 - 592. Error: 200000437 Image installation failed for image (for some Exam/Series/Image number) Open Long Description for full error varaiations Sat Jul 16 13:11:02 2005 Error: 200000437 Host: hsa2 Process: dms File: InstallImage.cxx Line: 316 Image installation failed for image 7297/1/1. Reason: Response from SdC timeout. Thu Feb 2 13:32:10 2006 Error: 200000437 Host: ct01 Process: dms File: InstallImage.cxx Line: 713 Image installation failed for image 2336/999/1. Reason: DMS internal error occurred. image path = %s Error code: 200000423 Error: 230009067 Retry queue entry See Long Description for error example. Tue Oct 19 09:24:04 2004 Error: 230009067 Host: dgct Process: queuemgr File: qmgr_list_mgmt.c Line: 2217 Function: Image Generation : SBC Processing Job Type: Prospective Helical Scan: 10989/2/1 Image: 10989/2/1 Retry queue entry Error: 200502002 Cannot allocate image space as one chunk. See long description for error example Error
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 101I.2_HSQXIM5 See LD for ClassM & C 101I.2_HSQXIM5 used on HiSpeed (YMS) only Class M software - 2395173 or 2401759 (6 pack) Class C software - 2395172 or 2401759 (6 pack) Patch G.6 Mandatory FMI# 25363 (P/N 5136941)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.