Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2601327992297726013279350089

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260132799 The Following Missing Digital Interface Boards Indicate A Failure In The 48V Power Cable see long description for complete error Wed Dec 2 22:17:53 2009 Error: 260132799 Host: orp Process: orp_subsys File: IfbConfig.cxx Line: 86 Wed Dec 2 22:17:52 2009 Host : Das Ermes # : 260132799 Exception Class : Abort Severity : Secondary File : IfbConfig.cxx Line# : 86 Function : No System Function Reported Scan Type : Scout Scan: 22977/1/1 The Following Missing Digital Interface Boards Indicate A Failure In The 48V Power Cable/Connector J8 On The Low Channel Backplane. Digital Interface Boards: 15-22

Error: 260132793 Digital Modules Have Not Been Detected By The Data Acquisition System See Long Description for complete error. Mon Feb 6 09:42:29 2006 Error: 260132793 Host: orp Process: orp_subsys File: IfbManager.cxx Line: 318 Mon Feb 6 09:42:28 2006 Host : Das Ermes # : 260132793 Exception Class : Abort Severity : Primary File : IfbManager.cxx Line# : 318 Function : No System Function Reported Scan Type : Static Scan: 50089/1/1 The Following Digital Modules Have Not Been Detected By The Data Acquisition System. See Previous Messages From Last Reset For Detailed Fault Inform

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues