GE Healthcare · CTBrightSpeed26011821150018
BrightSpeed HPower, Delight, Velocity or MERI HPower
Symptom
unable to scan after x-ray tube replacement with inverter max error .
Error message
Error: 260118211 ( 60-0313 )Inverter max. ILR current detected. This is a hardware detection of maximum current in serial resonant circuit. Z6,Z7= 0h,0h See long description for complete error Mon Feb 13 17:22:21 2006 Error: 260118211 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Mon Feb 13 17:22:20 2006 Host : Jedi Ermes # : 260118211 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50018/1/4 (60-0313)Inverter max. ILR current detected. This is a hardware detection of maximum current in serial resonant circuit.
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2246223-2 HV TANK HEMIT JEDI 60DC Used on 5.X 60KW and BrightSpeed 4/8/16 slice HPower Scanner Elite-16, Edge-8, Excel-4 See Long Description for details 2246223-2 HV TANK HEMIT JEDI 60DC Used on 5.X 60KW and BrightSpeed 4/8/16 slice HPower Scanner Elite-16, Edge-8, Excel-4 Part #: 2246223-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HV TANK HEMIT JEDI 60DC (SPARE) FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required