GE Healthcare · CTHiSpeed18241425267
Banging, clicking, moaning, squeaking, grinding noise, or some other noise in the table as it lowers or raises.
Symptom
Banging, clicking, moaning, squeaking, grinding noise, or some other noise in the table as it lowers or raises. Banging, or clicking, moaning, or squeaking, or grinding noise, or some other noise in the table as it lowers or raises
Error message
Error: 182414, Current limit fault detected by hardware. See LD for complete error Error: 182414, Current limit fault detected by hardware Wed Feb 23 21:14:53 2005 Error: 182414 Host: ETC Process: Elevation File: emcCurLimit.c 1.20 11/13/92 Line: 70 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 25267/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 21:14:53:830 Log Series: 445 HW error: Current limit fault detected by hardware. Current Position 4946 (.1 mm) Elev Direction: 0 (0 = Up 1 = Down).
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2294322-3 (replace 2294322-2) : Table Elevation Linear Actuator, Zeus thru H16 Table. See long description for details. Part #: 2294322-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Actiator, Linear, Zeus thru H16 Table, Pinned, R4 Ball Screw Fillet FRU: Yes Status: Orderable Old FRU Compatible Yes Comments Per ECR2079949, 212008898, 5/19/09. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 2294322-3 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2079949 Part# 46-296561P1 converts to Part #: 2294322-2 in FEMC Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Actuator, Linear, Zeus thru H16 Table, Pinned FRU: Yes Status: Orderable Parts Offered as: Current: 2294322-2 Harvested: Not Available Exchange: Not Available Comments: Added per ECR ms13452 041210
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required