GE Healthcare · CTLightSpeed1144566814877214
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Multiple recon retries and suspended images during an exam. See long description for details Multiple recon retries and suspended images during an exam. To get the images to recon you would have to resume the queue or reboot the system for the images to recon.
Error message
raw_data - lost chains " cross-linked with another chain ( found in the iceConsole.log ) See LD for Error Details To View Scan Data Disk Errors in the iceConsole.log cd /usr/g/service/log tail -100 iceConsole.log | more (tail –500 would show the last 500 lines of this log) (use the space bar to page through this log) Look for any Scan Disk Errors as seen below: "/raw_data - lost chains " cross-linked with another chain "/raw_data - lost chains " cross-linked with another chain "/raw_data/ex50752.CT01.1144566814.877214/ov.s7a13" has illegal cluster in chain "/raw_data/ex50752.CT01.1144566814.877214/ov.s7a14" has illegal cluster in chain "/raw_data/ex50752.CT01.1144566814.877214/ov.s7a18" has illegal cluster in chain
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2225004-10 73 GB destroked to 18GB 15000 Used on Octane OC with 9Gb OS hard drive. See Long Description for more detail. This part is used on Octane Operators Consoles that have a 9Gb OS hard drive and need a destroked Scan Data Disk. See SN0711061b for jumper settings. Part #: 2225004-10 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73 GB destroked to 18GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required