LightSpeed 4.X - GRE
Symptom
Less image space available than normal. See long description for examples. Less image space available than normal, even after removing all exams in Image Works. Example: H3 or GOC1 console should have 20,000 (512 images) available on a clean system with an Octane cpu with two 9 GB disk drives. But with the replacement of the 9GB drives with 18 and 73GB drives, image space may increase up to 130,000 (512 images) depending on software. Example: GOC2 console should have 100,000 (512 images) available on a clean system with an HP-8000 cpu with one 18GB disk and one 73GB disk. Example: GOC3 or 4 console should have 200,000 (512 images) available on a clean system with an HP-8200 cpu with one 36GB disk and two 73GB disks. Data above is approximate, but can be used to determine if there is a problem with disk space.
Error message
Error: 200502003, Cannot allocate image space for XXX images, allocated XX images. See LD for complete error Mon Feb 13 16:36:28 2006 Error: 200502003 Host: ls16dn Process: retroRecon File: srx_sma_manage_image_space.c Line: 690 Function: No system function reported Cannot allocate image space for 148 images, allocated 7 images. Error: 210000441 Image space allocation failed. See LD for complete error Mon Mar 26 12:51:14 2007 Error: 210000441 Host: roct Process: dms File: Initialize.cxx Line: 856 Image space allocation failed. size 3863552 Error code: -1 Error: 210000434 IMAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data See Long Description for complete error Fri Apr 21 07:41:23 2006 Error: 210000434 Host: lightspeed Process: dms File: SaveContext.cxx Line: 184 I MAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data. Error: 200109012 Image Allocation Failed. Cannot build scan request. See Long Description for complete error Wed Mar 2 18:39:33 2005 Error: 200109012 Host: ct01 Process: scanRx File: UIRxToScanner.cxx Line: 1307 Function: Data Acquisition : OC Processing Image Allocation Failed. Cannot build scan request.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2368697-3 HEWLETT-PACKARD HP8000 Computer used in Global Recon Engine (GRE) CONSOLE 2368697-3 GRE CONSOLE HP8000 2368697-3 HP8000 COMPUTER USED IN GRE CONSOLE Part #: 2368697-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: LINUX DDR WORKST'N PS2 MOUSE--100 PC FRU: Yes Status: Orderable Comments: 3/4/05 (212001261 Shawn Fox) HOST COMPUTER (HP). THISIS THE GOC2 HP 8000 COMPUTER. ***EXTREMELY IMPORTANT: NEW SOFTWARE, S/W LOAD PROCEDURE AND S/W OPTION KEYS ARE REQUIRED IF THE HP8000 COMPUTER IS REPLACED. Please contact the OLC for assistance with new Option Keys
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required