GE Healthcare · CTLightSpeedVCT290009008
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 290009008 Timeout waiting for Image Set Complete from IG3. (could be IG1 or IG2) See LD for description Tue Nov 13 12:57:56 2007 Error: 290009008 Host: darc Process: recon control File: ImageSetRequest.cxx Line: 5427 Timeout waiting for Image Set Complete from IG3. Will reset the IG.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5255299-20 VDIG VCT Data Aquisition and Image Generator Computer used on GOC6 5255299-20 VDIG VCT Data Aquisition and Image Generator Computer used on GOC6 Part #: 5255299-20 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VDIG - VCT Data Aquisition and Image Generator Computer Assembly FRU: Yes Status: Orderable Part #: 5255299-20 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VDIG - VCT Data Aquisition and Image Generator Computer Assembly FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required