LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. See long description for pop up message picture. A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. After clicking OK Fast Cal completes. ATTENTION: The system detected a problem with the DAS/Detector sub-system. Image Quality will be compromised. Scanning will NOT be disabled, but it is critical that the user call GE SERVICE to rectify the problem. Press 'OK" to proceed at the risk of Image Artifacts

Error message

Error: 230007014 Uncorrectable detector channels have been found for scan See LD for an example of entire error message. Fri Oct 28 09:32:03 2005 Error: 230007014 Host: presbyplano Process: badchanmgr File: badchanmgr.cxx Line: 1194 Uncorrectable detector channels have been found for scan 50029/2/2

Error: 230007011 Density Check - Module (X) exceeds number of shorts allowed (X) = modual number See LD for complete error Mon Oct 31 07:03:37 2005 Error: 230007011 Host: cpcvct2p Process: badchanmgr File: DensityMap.cxx Line: 63 Density Check - Module 6 exceeds number of shorts allowed (151 > 64). Sat Aug 15 14:04:26 2009 Error: 230007011 Host: vct Process: badchanmgr File: DensityMap.cxx Line: 63 Density Check - Module 24 exceeds number of shorts allowed (320 10).

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5136400-3 FRDM Center Module Kit for VDAS. see long description. Part #: 5136400-3 VCT CENTER DIGITAL MODULE KIT for VDAS Part #: 5136400-3 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: VDAS CENTER DIGITAL MODULE KIT for VCT FRU: Yes Status: Orderable ** FE ASSIST ** This is single center slot (27,28,29,30,31) VDAS64 Detector Module for field replacement. Refer to Service Note SN05091210e or later revision for any special instructions