Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260133175260133171260133176

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

"Scanner Hardware Stopped Scan" with CAM or Filter errors

Error message

Error: 260133175 Aperture Homing Problem See long description for full error Mon Sep 29 07:35:24 2008 Error: 260133175 Host: orp Process: orp_subsys File: aperture_home.cxx Line: 290 Mon Sep 29 07:35:22 2008 Host : Collimator Ermes # : 260133175 Exception Class : Abort Severity : Primary File : aperture_home.cxx Line# : 290 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Aperture Homing Problem

Error : 260133171 An error occured attempting to home CAM A or B. See Long Description for Details Wed Jul 29 08:19:20 2009 Error:260133171 Host: orp Process: orp subsysFile: aperture home.cxx Line: 353 Wed Jul 29 08:19:20 2009Host : Collimator Ermes # : 260133171 Exception Class : Abort Severity : PrimaryFile : aperture home.cxx Line# : 353 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 An error occured attempting to home CAM B.

Error: 260133176 Did not detect an encoder index pulse when homing CAM A / B. See LD for more details Did not detect an encoder index pulse when homing CAM A / B. Fri Jun 9 08:41:23 2006 Error: 260133176 Host: orp Process: orp_subsys File: aperture_home.cxx Line: 721 Fri Jun 9 08:41:22 2006 Host : Collimator

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5271284 VCT Collimator Motor and Coupling Assemb See long description for details Part #: 5271284 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: VCT Collimator Motor and Coupling Assembly FRU: Yes Status: Orderable Comments: Per ECR 2052055, 212008898, 4/7/08.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.