Pular para o conteúdo principal
GE Healthcare · CT23000650611433

Lightspeed 1.X - Octane

Symptom

Image reconstruction failures. Image reconstruction failures

Error message

Error: 230006506 Reconmgr received job_status error RECON_JOB_FAILED_RETRY from SDC Control See LD for complete error Sat Jul 1 19:39:27 2006 Error: 230006506 Host: ice Process: reconmgr File: rm_process_job.c Line: 491 Function: Image Generation : SBC Processing Job Type: Prospective Scout (s Scan: 11433/1/2 Image: 11433/1/2 Recon Seq Id: 2824 Reconmgr received job_status error RECON_JOB_FAILED_RETRY from SDC Control in the sdc done event.

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2216467 Pegasus Image Generator used on LS 1.x-4.x Octane, GOC1 and GOC2. Not GRE See long description for details Part #: 2216467 Part Type: GP REPAIRABLES Part Weight: 4.41 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: FAST IMAGE GENERATOR FRU: Yes Status: Orderable Comments: FLS - PEGASUS IMAGE GENERATOR PWA CIRCUIT BOARD. USED ON HELIOS 2. MS26561 DSM 2/24/01 PCN 220318 ERRONEOUSLY CHANGEDBOARD VERSION FROM E TO G (REVISION G IS WRONG, NOTHING SHOULD BE MANUFACTURED UNDER THAT REVISION, PER PCN 220318). ADDED BY MS20544, CSV ON 1/21/02. 7/19/02- C2R COMPLETED. MS12464 4/21/03 - FEILD RESTOCK FLAG RESET TO Y. MS12575 LIST PRICE $26371.00 GELS - SECOND OPINION PROCESS TO ASK OLC IN BUC BEFORE TO SHIP INFO FROM P DE LA FUENTE /14/06/TS GA0312 EN ATTENTE DE REGUL STOCK CONTACTER GILLES ANCELOT SH CSO 0300690 ITALIE I 24810431002,PCB PEGASUS IMAGE GENERATOR PWA See example below for GOC1/GOC2 ICE box with pegasus IG board.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.