VG/Infinia Acquisition could not be started after gantry moved to home position
Symptom
VG/Infinia Acquisition could not be started after gantry moved to home position After selecting 'Home' position, acquisition could not be started, GDU did not display "start" message, . The protocol acutal table height is 78mm, but GDU display table height showed 77mm. Acquisition could not be started because table did not move to home position. Afterr gantry reset GDU displayed 32# error.
Error message
Error #32 (Error 32) ZERO SWITCH DETECTED MISPLACED, AXIS=$ Axis # Axis Name 0 Lateral 1 Radial #1 2 Radial #2 3 N/A 4 N/A 5 Swivel #1 6 Swivel #2 7 Rotation 8 Linear 9 Height
Applies to
- Models
- Infinia
- Product
- VGH VG and VG Hawkey and Varicam Infinia I Infinia I Acquisition System Infinia II Infinia II Acquisition System Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software)
- Component
- *Patient Table *Patient Table *Patient Table 2199559 Zero Flag film 2199559 Zero Flag film
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- After software upgrade FMI40786, the Infinia application does not automatically start on boot up
- I need the product data sheet on catalog E4500AK x-ray in use warning light interface control panel for Infinia
- Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
- Where do I obtain the Non-Linear Correction tables for a VG/Infinia Hawkeye system if can't find the CD that came with the DAS assembly?
- Hawkeye CT scan Ring (Circular) artifacts in X-ray images appearing in the same radius in all of the images.