GE Healthcare · CTHiSpeed200002352200002357
HiSpeed Dual, ProSpeed FII Series
Symptom
The system will scan fine but as soon as you hit the End exam the system crashes and shuts down The system will scan fine but as soon as you hit the End exam the system crashes and shuts down.
Error message
Error: 200002352 A HOST_SHUTDOWN was initiatied by the unexpected termination of scanRx. See long description for complete error Error: Fri Jan 19 16:02:31 2007 Error: 200002352 Host: act8 Process: cupMonitor File: ssProcStop.c Line: 766 A HOST_SHUTDOWN was initiatied by the unexpected termination of scanRx. Mon Nov 28 10:52:30 2005 Error: 200002352 Host: kmcct2 Process: cupMonitor File: ssProcStop.c Line: 1065 A HOST_SHUTDOWN was initiatied by the death of scanRx. Error: 200002357 A HOST SHUTDOWN was initiatied by the unexpected termination of dms since other processes depend on it to come up. See long description for a sample error. Tue Mar 18 08:58:45 2008 Error: 200002357 Host: jghls64 Process: cupMonitor File: ssProcStop.c Line: 954 A HOST SHUTDOWN was initiatied by the unexpected termination of dms since other processes depend on it to come up.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed Dual, ProSpeed FII Series HiSpeed Dual, ProSpeed FII Series Product Name: Hispeed Dual ( Linux ) CT/Be2 A list of types. (a Product/hardware table) This console has an HP8000, HP8200, or HP8400 Computer (not GRE console) An On-site / Remote identification aid . (Example: check_config) HP Host computer A Table of options (included, optional, new introductions, …) HiSpeed Dual for China and Global B71402CA has the following as standard Options 1.5 sec fast scan, Power 200ma, Thichness 1mm, Raw data 5000, Image 100000 A list of Software that has run on this product (Product / Software Compatibility Table) With HP8000/8200 : • OS Version 4.5.15144980•Software Version 2.235165193 (except China)•Software Version 2.235172510 (for China only)•Patch 2.245179238•Patch 2.255197828•Patch 2.275267772 With HP8400 : •OS Version 6.2.95265475•Software Version 3.105270457•Patch 3.305267336•DVD Interchange 6.15270187•Patch 3.40 (See Note 8)5316222•DVD Interchange 7.05316345N/ Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links: Drawing Print 2320306-2CCT : http://gemspdm.med.ge.com/ematrix/common/eLibDocument.jsp?type=type_DrawingPrint&name=2320306-2CCT&rev=9 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- Patch V/R 3.40 / 3.41 / 3.50 For CT/Be-2 (HiSpeed Dual, ProSpeed FII) Patch's installed on V 3.30 HiSpeed Dual / Prospeed FII Systems (Be2) running software package V/R 3.10 + patch V/R 3.30 for Linux OC See long description for details Refer to Service Note GEHW SN110802A , which can be found on the Distributed Technical CDROM/DVD, or online at: Beijing CT Tech Pubs websites Software Version: Be2 V/R 3.40 / 3.41 / 3.50 Software Patch V/R 3.50 released as FMI 22825 Kit (P/N: 5367734) Part Numbers: The complete software set must be ordered 5148155-7 CT Be2 OS and App SW ASSY Consisting of the following software 5270457 Refresh Be2 Application Software VR3.10 5267336 Be2 Patch V/R3.30 for VR3.10 Application 5316222 Be2 Patch V/R 3.40 for 3.30 5322308 Be2 Patch V/R3.41 for 3.40 5316345 DVD Interchange Ver 7.0 Changes: Refer to the Service Note attached Compatibility & Description: "5148155-7 is Be2 Software Assembly of : OS6.5.1+3.10(Application)+3.30(Patch)+3.40(Patch)+DVD interchange (7.00). Impact: Software assemblies contain rating plate for site software version tracking, please archiving the rating plate when finishing S/W upgrading. Expected FMI: FMI 22820 kits (P/N: 5318410) for Be2 systems with HP8200PC/HP8400PC FMI22823 kits (P/N: 5330827) for Be2 systems with HP8000PC The 2 FMIs are Mandatory On Request. Expected Implementation Date: 06/13/2008
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required