Pular para o conteúdo principal
GE Healthcare · CTLightSpeed182900504121831825000144858

LightSpeed 4.X - GRE

Symptom

"Xray and drives power is disabled, please enable" message seen after rebooting. See long description for details. Message seen after rebooting, in the message area or in a popup message saying: " Xray and drives power disabled, please enable. "

Error message

Error: 182900 Stopping scan because a tableside emergency off button has been pressed Open Long Description to see example error Sun Jul 18 10:02:54 2004 Error: 182900 Host: ETC Process: Table Controls File: tc_scan.c 1.14 1/11/94 Line: 235 Function: No System Function Reported Scan: 50412/1/3 Type: Axial Exception Level: Pri/Most Severe Time: 10:02:54:710 Log Series: 57 Stopping scan because a tableside emergency off button has been pressed

Error: 183182 I/O board pilot relays are closed. PDU pilot relay not energized. HVDC Rail Voltage is xxxx See Long Description for more details. Error: 183182 I/O Bd. pilot relays are closed. PDU pilot relay not energized. HVDC is xxxx. ** Note the HVDC Rail Voltage. It is important in troubleshooting Tue Sep 21 06:13:59 2004 Error: 183182 Host: OBC Process: Generator Function: SYSTEM POWER CONTROL: HVDC Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe I/O board pilot relays are closed. PDU pilot relay not energized BCSTAT = 0 HVDC Rail voltage = 0V Sun Dec 12 18:11:57 2004 Error: 183182 Host: OBC Process: Generator File: gen_rotor.c 1.39 Line: 488 Function: SYSTEM POWER CONTROL : HVDC Control Scan: 44858/1/1 Type: Scout Ex

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-297093P1 AC/DC POWER SWITCH - used on Zeus tables See LD for more info 46-297093P1 HSA-CTI AC/DC POWER SWITCH Part #: 46-297093P1 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: AC/DC POWER SWITCH FRU: Yes Status: Orderable FLS - IBS PART AC/DC TOGGLE SWITCH Comments: FLS - IBS PART AC/DC TOGGLE SWITCH
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.