GE Healthcare · Nuclear MedicineInfinia5372534
ITF 5372534-1EN Hawkeye TTH Low Level FRU Replacement Procedure
Symptom
ITF 5372534-1EN Hawkeye TTH Low Level FRU Replacement Procedure Replace PCB or Tank Assemblies, thereby reducing the TTH Hawkeye1 / Hawkeye4 Material Service Cost. All TTH Hawkeye1 and Hawkeye4 subsystems of Infinia and VG Systems
Applies to
- Models
- Infinia
- Product
- Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software) Infinia II Infinia II Acquisition System Infinia I Infinia I Acquisition System VGH VG and VG Hawkey and Varicam
- Software
- Infinia II Hawkeye 4 & GP 2.105.030.* (SP2) Infinia II Hawkeye 4 & GP 2.105.030.* (SP2) See SN43847 : P/N 5389346 Software kit - Infinia II GP & Hawkeye 4 SP2 SW Release 2.105.030.3 See Infinia CIS 5191823-100 Infinia II + Hawkeye 4 Imaging System Software version 2.105.030.0 See SN43877 : P/N5396109 Standalone Patch only - Infinia and Hawkeye4 SP2 Cumulative patch Ver. 2.105.030.5 P/N5395606 Software kit - Infinia and Hawkeye4 SP2 and eManual for spare part 2.105.030.5 See FMI40828 : (Suspended) P/N 5400788 Infinia and HWK4 SP2 2.105.030.0 and Cumulative Patch 2.105.030.6 See ITF 5410176-1EN P/N 5410198 Infinia and Hawkeye 4 SP2 Accumulative patch Ver. 2.105.030.9 See SN43914 ( FMI 40834 ) ITF 5419572-1EN Infinia Hwk4 SP2 Cumulative Patch Installation 2.105.030.10 P/N 5419735 Infinia II GP & Hawkeye 4 SP2 Cumulative Patch Ver. 2.105.030.10 See ITF 5430930-1EN Infinia II GP & Hawkeye 4 SP2 Cumulative Patch 2.105.030.11
- Component
- *X-Ray (Generator, DAS, Controller) *X-Ray (Generator, DAS, Controller) 5372535 TTH HAWKEYE1 FRU TANK ASSEMBLY 5372535 TTH HAWKEYE1 FRU TANK ASSEMBLY 5372536 FRU PCB ASSEMBLY FOR TTH HAWKEYE1 AND HAWKEYE4 5372536 FRU PCB ASSEMBLY FOR TTH HAWKEYE1 AND HAWKEYE4 5372537 TTH HAWKEYE4 FRU TANK ASSEMBLY 5372537 TTH HAWKEYE4 FRU TANK ASSEMBLY
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- After software upgrade FMI40786, the Infinia application does not automatically start on boot up
- I need the product data sheet on catalog E4500AK x-ray in use warning light interface control panel for Infinia
- Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
- Where do I obtain the Non-Linear Correction tables for a VG/Infinia Hawkeye system if can't find the CD that came with the DAS assembly?
- Hawkeye CT scan Ring (Circular) artifacts in X-ray images appearing in the same radius in all of the images.