GE Healthcare · CTLightSpeed230023046230017082394680
LightSpeed RT16 & RT16 Xtra and Vision
Symptom
Fastcal is failing with beam obstructed errors. Fastcal is failing with beam obstructed errors
Error message
Error: 230023046 Calibration limit check failed. Beam may be obstructed See LD for complete error Tue Mar 22 20:12:04 2005 Error: 230023046 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3953 Calibration limit check failed. Beam may be obstructed Error: 230017082 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.3 The row number can be VERY confusing. Please see Long Description for how to decode the ROW NUMBER Tue Mar 22 20:12:04 2005 Error: 230017082 Host: ct01 Process: accs File: aircal_processing.c Line: 917 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.394680 Click Here to Decode the ROW NUMBER
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2354053 WATSON 16 DETECTOR 2354053 WATSON 16 DETECTOR 2354053 WATSON 16 DETECTORPart #: 2354053Part Type: GP REPAIRABLES Part Weight: 0.00 LBRRepairable: YesReturnable: YesHazardous Material: NDescription: WATSON 16 DETECTORFRU: YesStatus: OrderableComments: FLS - WATSON16 DET ASM W/DIODE UPGRADE. TO BE USED ON FORWARD PRODUCTION OF H16 SYSTEMS, PERCT PCN 220726. COLLECTION FOR C2R PER B CORTRIGHT. MS12464
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.