Symptom

Error 50-0204 Filament permanent open circuit Intermittent Error Code 50-0204 (Filament permanent open circuit).

Error message

50-0204 Jedi error code See long description for details Filament permanent open circuit - Issued by the Heater board when an open has been detected and 3 restarts have been tried without success within 100 ms. Potential causes: 1. X-ray tube filament open 2. Heater to HV Tank cable 3. Cathode HV cable or pin contacts 4. Open circuit in filament transformer inside HV tank.

Applies to

Models
Brivo, Definium, Precision, Optima, Discovery
Product
Brivo DR-F Definium 8000 Precision 500D Optima XR640 Discovery XR650 PROTEUS XRa
Area
X-ray R&F