GE Healthcare · CTHiSpeed219800
HiSpeed QX/i - Octane
Symptom
Unable to scan or perform tube warmup When the tube rotor spins up or down it may sound like a jet plane is landing and the gantry covers vibrate
Error message
Error: 219800 The HEMRC AC Drive reported fault code: F5 - Overvoltage Fault See longdescription for full error Fri Mar 30 11:14:04 2004 Error: 219800 Suite: CT01 Host: OBC Process: Generator File: HEMRClogFault.c 1.2 Method: No Method Line: 196 Function: XRAY GENERATION : Tube Rotor Control Scan: 0/0/1 Type: Static Exception Level: Secondary Time: 10:14:01:725 Log Series: 29 The HEMRC AC Drive reported fault code: F5 - Overvoltage Fault The Drive internal DC bus voltage has exceeded 810V. Possible failure of HEMRC I/F Board Chopper Control or excessive motor regeneration from X-ray tube during braking.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-313346P1 60 amp Fuse in CPDU. 60 amp Fuse Part# 46-313346P1 in CPDU. 60 amp Fuse Part# 46-313346P1 in CPDU. Part #: 46-313346P1 Part Type: GP CONSUMABLES Part Weight: 0.14 LBR Repairable: No Returnable: No Hazardous Material: N Description: 060.00 A 0600V CARTRIDGE SLOW UL/C FRU: Yes Status: Orderable Comments: FLS - CLASS 1 FRU PDV PER MFG 7-92
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required