GE Healthcare · CTLightSpeed26010720650180260107204
LightSpeed 5.X (80/100kW)
Symptom
Gantry does not rotate. Customer noticed that gantry does not rotate.
Error message
Error: 260107206 Axial Motor Control (AMC) logic will NOT accept commands until the Axial Drive is properly configured/initialized. See Long Description for full error Wed Apr 7 21:04:02 2004 Error: 260107206 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 3771 Wed Apr 7 21:03:47 2004 Host : Table Gantry Processor Ermes # : 260107206 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 3771 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Axial Motor Control (AMC) logic will NOT accept commands until the Axial Drive is properly configured/initialized. (Cmd: 27 Current State:0 Flags:0x8 ReInit Status:3). or Host : Table Gantry Processor Ermes # : 260107206 Exception Class : Abort Severity : Secondary File : amc mgr.c Line# : 3883 Function : No System Function Reported Scan Type : Axial Scan: 50180/1/2 Axial Motor Control (AMC) logic will NOT accept commands until the Axial Drive is properly configured/initialized. (Cmd: 21 Current State:0 Flags:0x8 ReInit Status:3) Error: 260107204 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration. Reset scan hardware to retry. See Long Descript
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P101 - FUSE 20 AMPS 700 VOLTS - Fuse used in Axial Dynamic Brake and HEMRC (HSA - 1.X) and also in HEMRC (2.X-5.X) See LD for some of the locations of this fuse 46-170021P101 - FUSE 20 AMPS 700 VOLTS many pictures below 46-170021P101 - FUSE 20 AMPS 700 VOLTS - Fuse for Axial Dynamic Brake Part #: 46-170021P101 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE 20 AMPS 700 VOLTS FRU: Yes Status: Orderable Comments: FLS - FUSE 20 AMPS 700 VOLTS USED ON FAST RHAPSODE GWENN 08992 7/2/97 Fuse for 5.X Axial Dynamic Brake part found in LIGHTSPEED 5.X ILLUSTRATED PARTS LIST DIRECTION 2349187-100, REVISION 14 Chapter 4 Gantry (Pro 16) Section 1 HP100 Vertical Assembly Table 4-17 Fuse for PRE 5.X Axial Dynamic Brake Fuse for HEMRC Rotor controler (HSA-1.X) Fuse for HEMRC Rotor Control (2.x-4.x)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.