Pular para o conteúdo principal
GE Healthcare · CTLightSpeed23000006750000

LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2

Symptom

Scanner Hardware Stopped Scan with DIP and view length errors. Scanner Hardware Stopped Scan with DIP and view length errors

Error message

Error: 230000067 View Length error See LD for more details Tue Aug 22 12:35:04 2006 Error: 230000067 Host: ct01_sbc Process: dataacq File: dip_if.cxx Line: 416 Function: Data Acquisition : Scan Data Acquisition Scan Type: Axial Scan: 50000/1/4 Scan Seq Id: 4000 Unexpected event occurred during scan data transfer. Event bit: 5 Count: 1 Description: View length error Scan aborted.

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2293408 fiber optic cable from the DAS DCB to the RF Transmitter on the gantry rotate frame See long description for details Part #: 2293408 Part Type: GP CONSUMABLES Part Weight: 0.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: CABLE FIBER OPTIC FRU: Yes Status: Orderable FLS - CABLE FIBER OPTIC, 2840 MM +/- 20MM. USED ON WARP3.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.