Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2100004072005200152000004164294967295200000415

LightSpeed 5.X (60kW)

Symptom

Will not scan with "Not enough image space exists. Please remove images". See Long Description for screen capture. Scanner would not allow scouts, axial or helical scanning with message on left screen: Not enough image space exists. Please remove images.

Error message

Error: 210000407 Image space query failed. See Long Description for more details Mon Feb 27 07:19:00 2006 Error: 210000407 Host: csbipct2 Process: dms File: Initialize.cxx Line: 1085 Image space query failed. Error code: -1

Error: 200520015 Updating image space, received a bad status code: See the long description for different status codes Sat Jan 22 18:23:06 2005 Error: 200520015 Host: chr09bc1cti Process: environmentMgr File: ImageSpaceFormMgr.c Line: 117 Updating image space, received a bad status code: 200000416 Reason: Response from SdC timeout you may also see this error with different status codes, such as Fri Jul 28 05:12:48 2006 Error: 200520015 Host: vct1 Process: environmentMgr File: ImageSpaceFormMgr.c Line: 121 Updating image space, received a bad status code: -1 Reason: OS error : -1 --- Unknown error 4294967295 Wed Nov 16 08:23:54 2005 Error: 200520015 Host: gect_qxi Process: environmentMgr File: ImageSpaceFormMgr.c Line: 121 Updating image space, received a bad status code: 200000415 Reason: Invalid data from SdC.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5117866-51 Host Computer, CTPET HP8200 with FX1500 Graphics Card and 4GB Memory - replaces -24, -28, and -48 versions See Long Discription for Details Part #: 5117866-51 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Host Computer, CTPET HP8200 with FX1500 Graphics Card and 4GB Memory FRU: Yes Status: Orderable Comments: Per ECO 2042338, 5117866-51 is a substitute for 5117866-28 for Brightspeed systems. 212008898, 6/13/07. ** FE Assist ** For VCT systems must have software release 07MW18.4, p/n 5213049 (deployed by FMI 25394, p/n 5195244), and perform a LFC to support this FRU if system is not currently running sw rls 07MW18.4. or newer. For other CT systems , must order sw releases 07MW11.10, p/n 5220443 (deployed by FMI 25389, p/n 5191771)and perform a LFC to support this FRU if system is not currently running the sw release, or newer (9/12/07 212001261) 5115333 VCT GOC5 HOST HP 8000 computer New part number 5117866-28 Description: Host Computer, CT USA VCT XW8200 Part #: 5117866-28 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Host Computer, CTPET HP8200 with FX1400 Graphics Card and 4GB Memory FRU: Yes Status: Orderable Comments: Added per VCT RSPL, 040522 ms13452 This part is direct replacement for 5115333 . ms13452 Part #: 5117866-24 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 60.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Host Computer, CT USA 4,8,16-slice XW8200 FRU: Yes Status: Orderable Comments: Added per pathfinder RSPL, this part is not replacement validated for xw8000. ms13452 *** 2ND OPINION NOTES *** This part has a High NDF ( No Defect Found ) rate. Attempt to reload OS and Application software before replacement. ( 212024490 - 12/20/2006 )
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.