GE Healthcare · CTBrightSpeed2601182155036826011821410132
BrightSpeed Select, Delight Select or MERI Select
Symptom
Scanner Hardware Fault with high voltage spits. Scanner Hardware Fault with high voltage spits
Error message
Error: 260118215 (60-0318) kV did not reach 75 percent after 20ms See Long Description for full error Sat Mar 25 15:35:28 2006 Error: 260118215 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Sat Mar 25 15:35:27 2006 Host : Jedi Ermes # : 260118215 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50368/1/3 (60-0318)kV did not reach 75 percent after 20ms. Indicates that there was no HV ON signal generated for exposure time count-up. Error: 260118214 (60-0317) Spit Ratio error See Long Description for complete error. Fri Dec 16 23:26:55 2005 Error: 260118214 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Fri Dec 16 23:26:54 2005 Host : Jedi Ermes # : 260118214 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Axial Scan: 10132/2/1 (60-0317)Spit Ratio error. PPC board has detected the rate of tube spits during exposure has reached the limit (see theory of operation, software section).^C Z6,Z7= 4h,0h
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- D3146T SOLARIX 350 TUBE UNIT USED on BrightSpeed Select (lite) See long description for details Part #: D3146T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: REPLACEMENT SOLARIX 350 TUBE UNIT FRU: Yes Status: Orderable Comments: red tube, added per to BrightSpeed Lite RSPL by sso#305010201
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required