GE Healthcare · CTLightSpeedVCT260111601260111603
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Step and Shoot (SmartStep or SmartView) hand controller not functioning See LD for more Step and Shoot (SmartStep or SmartView) hand controller not functioning When customer attempts to use the hand controller for VCT step and Shoot, the system will not initialize into step and shoot mode.
Error message
Error: 260111601 IHHC reset timeout. See LD for complete error Thu Oct 20 12:41:32 2005 Host : Table Gantry Processor Ermes # : 260111601 Exception Class : Abort Severity : Primary File : hc_exec.c Line# : 766 Function : No System Function Reported Scan Type : None/Unknown Scan: 4628/4/0 IHHC reset timeout. Error: 260111603 IHHC watchdog timeout. See long description for a full sample error. Wed Jul 2 11:44:37 2008 Error: 260111603 Host: tgp Process: tgp subsys File: hc wdog.c Line: 230 Wed Jul 2 11:44:36 2008 Host : Table Gantry Processor Ermes # : 260111603 Exception Class : Abort Severity : Primary File : hc wdog.c Line# : 230 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 IHHC watchdog timeout.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 2199945-2 CT SONATA FOOTSWITCH FOR 2X/3X/4X See long Description for details Part #: 2199945-2 Part Type: GP CONSUMABLES Part Weight: 1.24 LBR Repairable: No Returnable: No Hazardous Material: N Description: CT SONATA FOOTSWITCH FOR H2 FRU: Yes Status: Orderable Comments: FLS - CT SONATA FOOTSWITCHFOR H2. THE CABLE LENGTH IS LONGER THAN 2199945. THE -2 P/N IS BACKWARD COMPATIBLE, BUT IT ISBEING CREATED FOR A NEW PRODUCT OFFERING, PER PCN 220511. MS20544 ON 5/24/02.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.