Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2001100262002880092900090021183997163967569200270032

LightSpeed 4.X - GRE

Symptom

System takes a scout but the tech cannot see it in ExamRx - document in RFS - critical issue System takes a scout but the tech cannot see it in ExamRx.

Error message

Error: 200110026 OC Processing Can not find scout scans See long description for full error Sun Dec 23 16:25:29 2007 Error: 200110026 Host: vct1 Process: scanRx File: ViewEdit.cxx Line: 9807 Function: Data Acquisition : OC Processing Can not find scout scans. AutoMa will be disabled for Series:3

Error: 200288009 ScanControl See Long Description for more detail on this error Sat Jan 22 08:39:24 2005 Error: 200288009 Host: nph1 Process: afQueueMgr File: /gspnet/blr/component/commondisplay/view/ Line: 2112 ScanControl::askForScoutImagePaths() Could not acquire image path(s) for xr scout. Thu Aug 24 01:43:55 2006 Error: 200288009 Host: klsp Process: afQueueMgr File: /data/blr/component/commondisplay/view/cd Line: 3512 ScanControl::printImageListInfo() The disk is 99 percent full!. Can't write to directory /usr/g/AfRecovery Tue Apr 21 09:01:47 2009 Error: 200288009 Host: SLMCHICT Process: afQueueMgr File: /gspnet/blr/component/commondisplay/view/ Line: 5366 ScanControl::unlockEverythingInTheSeries() Error: lockSeries() failed for exam: 7160 series: 4 path: /usr/g/sdc_image_pool/images/p81/e81/s370/i18722.CTDC.1

Error: 290009002 Error failed in opening file See long description for ful

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues